Ambient Temperature Range
- Operating & Storage: -10° to 60° C (14° to 140° F)
Approvals & Certifications
- Calibration Certificates
- CE
- EU Directives
- FDA
- PED 2014/68/EU
- UL (C,US): UL 61010-1 + CAN/CSA C22.2 No.61010-1
Circuit Protection
- Short Circuit: Protected for Transistor Outputs
Consumption
- Current [3 Outputs Meter (2-Wire)]: ≤ 25 mA at 14 Vdc without Transistors Load, with Current Loop
- Current [4 Outputs Meter (3-Wire)]: ≤ 5 mA at 12 Vdc without Transistors Load, without Current Loop
- Current: ≤ 1 A with the 2 Transistors Loads
Diameter
- Pipe: 0.9 to 4.3” (25 to 110 mm)
Electrical Connectors
- 3 Outputs Meter (2-Wire): 1 x 5-Pin M12 Male Fixed Connector
- 4 Outputs Meter (3-Wire): 1 x 5-Pin M12 Male + 1 x 5-Pin M12 Female Fixed Connectors
- Shielded Cable
Enclosure, Body Material
- 316L Stainless Steel, 1.4404
- PPS
Humidity
- ≤ 85 % Relative Humidity, Non-Condensing
Loop Impedance
- 1,100 ohms at 36 Vdc, Max.
- 3 Outputs Meter (2-Wire): 180 ohms at 14 Vdc, Max.
- 4 Outputs Meter (3-Wire): 100 ohms at 12 Vdc, Max.
- 610 ohms at 24 Vdc, Max.
Materials of Construction
- Display: Polycarbonate (PC)
- Fixed Connector Mounting Plate: 316L Stainless Steel, 1.4404
- Fixed Connector: Brass Nickel-Plated
- Navigation Key: PBT
Measuring Error
- Conductivity Measurement Deviation: ± 3 % of Measured Value
- Temperature Measurement Deviation: ± 1° C (1.8° F)
Measuring Range
- Conductivity Measurement: 0.05 to 10,000 µS/cm
Output
- 4-20 mA
- Transistor (NPN): 0.2-36 Vdc
Output Format
- Open Collector (700 mA, 0.5 A) - Max.
Power Supply
- 3 Outputs Meter (2-Wire): 14-36 Vdc, Filtered & Regulated
- 4 Outputs Meter (3-Wire): 12-36, Filtered & Regulated
Pressure Range
- psi: 0 to 232 (kPa: 0 to 1,600, bar: 0 to 16) - Max.
Process Temperature
- With G 1-1/2” PVC Nut Connection: 0° to 50° C (32° to 122° F)
- With G 1-1/2” PVDF Nut Connection: -20° to 100° C (-4° to 212° F) Restricted by the used Adaptor
Resolution
- Conductivity Measurement: 1 NS/cm
- Temperature Measurement (Internal Resolution): 0.1° C (0.18° F)
Response Time
- 10 to 90 %: 150 milliseconds
Reverse Polarity Protection
- Protected Against Reverse Polarity
Sensor, Probe Material
- Conductivity Sensor: 316Ti Stainless Steel 1.4571
- Conductivity Sensor: PVDF
Temperature Range
- Minimal Temperature Range: 10° to 20° C (50° to 68° F) Corresponding to 4-20 mA
- Temperature Measurement: -20° to 100° C (-4° to 212° F)
Wetted Materials
- Electrode: 316Ti Stainless Steel 1.4571 for Cell Constant C = 0.01 or C = 0.1
- Electrode: Graphite for Cell Constant C = 1.0